ClearView Radiation Shielding – Pluvicto (Lu-177 / Lutetium) Administration Shield

ClearView Radiation Shielding is being used in diagnostic imaging and therapies in over 500 hospitals with various products for Theranostics (in use or under development) as an alternative to lead. 

Specifications

  • Shielding
    • Beta: 0.25″ Acrylic
    • Lu-177: 96%
    • 0.3″ Lead Eq.
    • Exposure with Shield: 0.78mR/min.

  • Shielding information for in-situ dose transport
    •  Exposure: 1 meter (3.3ft) from shield with 209 mCi Lu-177: < 1 mRem/hr
  • Specifications 20cc
    • Syringe Shield
      • O.D.: 3.5″, I.D.: 0.9″
      • Length: 3.75″
      • Weight: 1020g
    • Cradle Footprint
      • 5.5″H x 4.75″W x 5.0″ L
      • Weight: 975g
  • Specifications 12cc
    • Syringe Shield
      • O.D.: 3.25″, I.D.: 0.75″
      • Length: 2.85″
      • Weight: 850g
    • Cradle Footprint
      • 4.2″H x 4.0″W x 4.0″L
      • Weight: 800g
Description Part Number
20cc Syringe Shield Assembly R100186
12cc Syringe Shield Assembly R100202
**Check Syringe Compatibility

Material Properties

  • DOT (US), IATA, IMDG certified non-hazardous and environmentally friendly.

  • NFPA, HMIS rating 0 (Fire, Physical Contact and Reactivity)

  • Does not create internal sources of radiation.

  • Non-toxic, non-abrasive to human contact.

  • Contents safe as per OSHA and EHS regulations