ClearView Alpha Particle Syringe Shield
Alpha Particle Radioligand Therapies (RLTs) require shielding from
primary Alpha particles, but also gamma’s in the decay chain.
Radium’s ClearView Alpha Particle Syringe Shield is designed to
shield hospital staff from radiation sources in Alpha RLT’s and
provide reliable dose handling equipment which supports timely
patient care.
Part Numbers:
- R100151 for 20ml syringe for 20 MBq
doses - R100232 for 5ml syringe for 5 MBq doses
Shielding Analysis:
- Source: 20 MBq or 0.540 mCi Actinium
(Ac) 225 - Testing Method: Ludlum 44-9 with 14-C
pancake probe - Syringe Used: BD 20 ml syringe.
- Unshielded Exposure: 126.4 mR/hr
- Exposure on contact: 17mR/hr, 86.5%
attenuation - Exposure 1’: 0.9 mR/hr (1 ft. away)
For questions, please contact sales@radiuminc.com